Product Overview
Fixed Gas Detectors | Xgard
Single Gas & Multi-gas Detectors | Fixed Point Gas Detection & Portable Gas Safety Applications | Industrial Locations & Hazardous Area Zones
Crowcon Xgard | ATEX Zone 0, Zone 1 & Zone 2 Hazardous Areas
Crowcon Xgard is a versatile range of fixed gas detectors which meet gas detection specifications for a wide range of applications and industries worldwide. The Crowcon Xgard can be found in research facilities, water treatment, power, oil and gas and manufacturing facilities.
This range of gas detectors feature a number of sensors including electrochemical sensors for toxic gases and oxygen, poison resistant pellistors to detect flammable gases and thermal conductivity sensors to monitor the percentage volume of concentrations of gases.
The Crowcon Xgard gas detector has both intrinsically safe (Ex i) and flameproof (Ex d) options available; certifications include IECEx, ATEX, UL, GOST-R and MED.
A more cost effective version is the Crowcon XSafe which is suitable for use within non hazardous areas.
Crowcon Xgard Gas Detector – Product Range
The diagram below illustrates how to determine which gas detector best suits the requirements of the end user.
Key Features
- Low Cost – Xgard detectors are designed for easy installation and maintenance to keep costs down. The three junction box options are all designed to make replacement of sensors and sinters extremely simple and spare sensors simply plug in. Many spare parts are common to all Xgard models, which keeps spares holding requirements to a minimum.
- Wide Range of Sensors – Poison resistant pellistors, for all flammable detection needs including hydrocarbons, hydrogen, ammonia, jet fuel, leaded petrol and vapours containing halogens. Electrochemical sensors are used to detect a vast range of toxic gases and oxygen. Thermal conductivity sensors are available to monitor % volume concentrations of gases.
- Flexible Installation Options – Xgard gas detector is designed for either wall or ceiling mounting without the need for additional brackets. The Xgard can accommodates M20, 1/2″ NPT or 3/4″ NPT cable glands to suit all site requirements. High temperature models are available for hot environments (up to +150ºC). Accessories are available for duct mounting and sampling applications as well as remote gassing for simple sensor checking.
- Rugged & Reliable – Xgard is manufactured using a choice of three materials: glass reinforced nylon, highly durable aluminium with a tough polyester coating, or 316 stainless steel for ultimate corrosion resistance. All versions are designed to operate even in the harshest conditions. Spray deflectors and weatherproof caps are available for use in areas subject to regular wash-downs, or offshore environments. All models have been validated to the functional safety standard IEC 61508 (SIL 1 to SIL 3)
Crowcon Xgard Bump Testing & Calibration
Routine bump testing is recommended for the range of Crowcon fixed gas detectors prior to each use. Bump testing is a brief application of a gas to the sensor to invoke a prescribed reaction, ensuring the gas detector reacts as expected.
Bump testing provides a quick, easy and direct method of verifying gas detectors are performing as required and takes a lot less time than ‘zeroing’ the device.
The bump test is widely accepted as a suitable method of testing gas detection equipment.
Fixed Gas Detector Gases And Ranges Detected
*Ranges not available for Xsafe or Xgard Type 4 -5 LTEK & STEL figures are derived from the UK HSE document: EH40 2011. Alternative thresholds may apply in countries outside the UK LEL figures derived from EN60079-20-1 : 2010 *1 Current limits advised in UK *2 Nominal 0-100ppm range with carbon monoxide (CO)
CROWCON xgard FIXED GAS DETECTOR HAZARDOUS AREA ATEX CERTIFIED GAS DETECTOR
Hazardous Area Classification
Type 1 Detector
ATEX: II 1G Ex ia IIC T4 Ga (Tamb -40 to +55° C)
UL/cUL: Class I, Div 1 Groups A, B, C, D
IECEx GOST-R
Type 2 Detector
ATEX: II 2GD Ex d IIC T6 Gb (Tamb -40 to +50° C)
UL: Class I, Div 1 Groups B, C, D
IECEx GOST-R
Type 3 Detector
ATEX: II 2GD Ex d IIC T4 Gb (Tamb -40 to +80º C) Ex d IIC T6 Gb (Tamb -40 to +50° C) Ex tb IIIC T180º C Db
UL: Class I, Div 1 Groups B, C, D
IECEx GOST-R
Type 4 Detector
ATEX: II 2 GD Ex d IIC T3 Gb (Tamb -40 to +150º C)
Type 5 & Type 6 Detector
ATEX: II 2 GD Ex d IIC T6 Gb (Tamp -40 to +50º C) Ex d IIC T4 Gb (Tamb -40 to +80º C) Ex tb IIIC T180º C Db
UL: Class I, Div 1 Groups B, C, D
IECEx GOST-R
Visit hazardous area gas detectors for explosion and flameproof gas monitoring and detection instruments with certification for potentially explosive atmospheres, including Zone 1 and Zone 2 (according to ATEX and IECEx) and Class/Division categories (according to North American standards).