Sick MCS300P

Simultaneous process monitoring of up to 6 measuring components

The MCS300P is an extractive gas analyser manufactured by SICK which is used for the simultaneous process monitoring of up to six measuring components. It is an extractive process photometer for measuring gaseous or liquid media.

SICK MCS300P measures IR and VIS active components with variable measuring ranges from very low (ppm) to high (vol%) concentrations. For monitoring of toxic or flammable mixtures, it has special process cells with safety devices like twinseals and flushing gas feeds.

The heatable cells made of corrosion-resistant materials have a high pressure resistance.

Automatic adjustment, innovative operation concept and modern communication protocols make the SICK MCS300P an all-purpose photometer, even for potentially explosive atmospheres.

Typical Applications for SICK MCS300P Gas Analysers

  • Process monitoring in production of vinyl chloride or isocyanate
  • Process control in the chemical industry
  • Process monitoring in the production of plastics
  • Monitoring of exhaust gas purification plants
  • Determination of the water content of solvents

Features

  • Simultaneous measurement of up to 6 components
  • Process cells up to 60 bar and 200 °C
  • Automated sample point switching
  • Integrated adjustment unit (optional)
  • Safety devices for measurement of toxic or flammable mixtures
  • Extended operation via PC and SOPAS ET software
  • Flexible I/O module system
  • Automatic adjustment without expensive test gases
  • Integration into existing networks
  • Integration of external parameters like temperature or pressure
  • Suitable for explosive environments
  • Easy installation and maintenance thanks to transparent, compact design

Technical SPECIFICATION

Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.

Dimensional Drawings – sick mcs300p – Standard Version

All dimensions in mm (inch); all weights in kg.

All dimensions in mm (inch); all weights in kg.

Dimensional Drawings – sick mcs300p – Ex Version

All dimensions in mm (inch); all weights in kg.

All dimensions in mm (inch); all weights in kg.

FOR EXPLOSIVE ATMOSPHERES & Hazardous Areas

Key Product Categories: Control Panels | Plugs & Sockets | Isolators | Enclosures & Junction Boxes | Lighting | Control Stations | Motor Starters | Heat Trace Cables & Systems | Gas Detection & Detectors | Fire Detection & Detectors | Heat Detectors | Electrical Heating & Heaters

Also Process Instrumentation Products: Ashcroft Pressure Gauges | ASCO Valves | Katronic Flow Meters | KROHNE Flow Meters | VEGA Level Sensors | Rotronic Temperature & Humidity Sensors | SIKA Pressure Gauges

Technical Specifications

BrandSICK
CategoryGas Detection
Measured ValuesBr2, CCl4, C2Cl4, CCl2F2, C3F6, CH4, C2H2, C2H4, C2H6, C3H4, C3H6, C3H8, C4H6, C4H8, C4H10, C6H6, C6H14, C7H8, CH3Cl, CH2Cl2, CHCl3, C2H3Cl, C2H2Cl2, C2H4Cl2, C2HCl3, C6H5Cl, C6H4Cl2, CHCl2F, CH3OH, C3H7OH, C2H4(OH)2, CH3CHO, HCOOH, CH3COOC2H3, C2H2O, Cl2, CO, CO2, COCl2, HCl, HCN, HF, H2O, NH3, NO, N2O, NO2, R-NCO, SO2, SiF4, many other IR / VIS active gases and liquids
Measurement PrinciplesInterference filter correlation, Gas filter correlation
Spectral RangeVIS version: 300 nm … 1200 nm IR-version: 1200 nm … 12000 nm
Measuring RangesMore than 60 measuring components available (depending on concentration and sample gas composition) Up to 6 components simultaneously 2 measuring ranges per component Automatic measuring range switching (adjustable) 2 limit values per component Measuring ranges depend on application and combination of measuring components
Response Time (t90)30 s … 120 s adjustable
Zero Point DriftVIS version: < 1 %: of smallest measuring range per day IR-version: < 2 %: of smallest measuring range per week
Detection Limit< 2 %: relative to measuring range end value
Ambient Temperature+5 °C … +40 °C Temperature change maximum ±10 °C/h
Storage Temperature–20 °C … +60 °C
Ambient Humidity≤ 80 % Non-condensing
Hazardous Area ApprovalsIECEx Ex pxb IIC T3…T4 Gb Ex pzc IIC T3…T4 Gc Process cuvette: Ex eb IIC Gb ATEX II 2G Ex pxb IIC T4 Gb II 2G Ex pxb IIC T3 Gb Process cuvette: II 2G Ex e IIC Gb II 3G Ex pzc IIC T4 Gc II 3G Ex pzc IIC T3 Gc
Electrical SafetyCE
Enclosure RatingIP65: measuring cuvette depending on version
Analog Outputs2 outputs: 0/4 … 22 mA, + 500 Ω Electrically isolated; max. number of outputs depends on application
Analog Inputs2 inputs: 0/4 … 22 mA, 100 Ω Electrically isolated; max. number of inputs depends on application
Digital Outputs5 outputs: 2 power relays, electrically isolated; 3 outputs, potential-free; max. number of outputs depends on application
Digital Inputs4 inputs: Open contacts, potential-free; max. number of inputs depends on application
Communication InterfaceModbus, Ethernet Remark: Not in the Ex version Type of Fieldbus integration: TCP Function: Connection to SOPAS ET software or OPC server
Number1
IndicationStatus LEDs: “Power,” “Failure,” and “Maintenance request”, LC display
InputFunctional keys
OperationVia LC-display Via software SOPAS ET (not in the Ex-version)
Dimensions (W x H x D)Dimensions may vary. For details, see the dimensional drawings.
WeightSee dimensional drawings
Material In Contact With MediaMeasuring cuvette depending on version
MaterialAluminium, coated
Power SupplyVoltage: Standard version: 115 V / 230 V, Ex-version: 230 V, Ex control device: 230 V Frequency: Analyzer: 50 … 60 Hz, Ex control device: 48 … 62 Hz Power consumption: Analyser: ≤ 230 W, With cuvette heating: ≤ 805 W, With 2nd heating: ≤ 1,450 W
Corrective FunctionsCross-sensitivity compensation of up to 6 interferents
CuvetteLengthWeight
L1L2L4
FGK224 (8.82)1000 (39.37)494 (19.45)39.5
PGK190, FGK299 (11.77)1080 (42.52)569 (22.40)43
PGK20399 (15.71)1180 (46.46)669 (26.34)45
PGK50699 (27.52)1480 (58.27)969 (38.15)51
PGK75949 (37.36)1730 (68.11)1219 (47.99)56
CellLengthWeight
L1L2
FGK168 … 229100033.5
PGK190, FGK299108037
PGK20399118039
PGK50699148045
PGK75949173050
CuvetteLengthWeight
L1L2L4
PGK10 Ex299 (11.77)1080 (42.52)569 (22.40)43
PGK20 Ex399 (15.71)1180 (46.46)669 (26.34)45
PGK50 Ex699 (27.52)1480 (58.27)969 (38.15)51
PGK75 Ex949 (37.36)1730 (68.11)1219 (47.99)56
CellLengthWeight
L1L2
PGK10 Ex299 (11.77)1080 (42.52)37
PGK20 Ex399 (15.71)1180 (46.46)39
PGK50 Ex699 (27.52)1480 (58.27)45
PGK75 Ex949 (37.36)1730 (68.11)50