SICK GME700 | Extractive Gas Analyser For Process & Emission Monitoring

Sophisticated process analysis “brought into line”

BrandSICK
CategoryGas Detection
  • Quoted to your quantity
  • Documentation on request
  • International delivery

Product Overview

SICK GMe700

Sophisticated process analysis “brought into line”

The GME700 is an extractive gas analyser manufactured by SICK which can measure over 60 different gas compounds and is used for process and emission monitoring.

Reliability and precision are features of the SICK GME700 laser gas analyser. Based on the principle of TDLS (Tunable Diode Laser Spectroscopy) and by using specific light absorption, the GME700 is able to measure several gas components, such as NH3, HF, or HCl.

The compact 19″ unit includes the analyser and evaluation unit as well as a heated multipath measuring cell.

Typical Applications for SICK GME700 Gas Analysers

  • Ammonia slip measurement in DeNOx plants
  • Hydrogen chloride monitoring in raw gas before scrubber in waste incineration plants
  • Measurement of hydrogen fluoride in semiconductor production
  • Ammonia measurement in urea production or composting plants
  • Plastics processing
  • Ammonia measurement in exhaust gas of combustion engines

Features

  • High selectivity due to high spectral resolution
  • No calibration required
  • No moving parts: minimal wear and tear
  • Heated multipath measurement cell
  • Hot-wet measurement technology
  • Optimal adaptation to measuring task through suitable gas conditioner
  • High reliability during operation
  • Can also be used in harsh ambient conditions
  • Easy integration into existing extractive systems
  • No problematic test gases required

Technical SPECIFICATION

Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.

FOR EXPLOSIVE ATMOSPHERES & Hazardous Areas

Key Product Categories: Control Panels | Plugs & Sockets | Isolators | Enclosures & Junction Boxes | Lighting | Control Stations | Motor Starters | Heat Trace Cables & Systems | Gas Detection & Detectors | Fire Detection & Detectors | Heat Detectors | Electrical Heating & Heaters

Also Process Instrumentation Products: Ashcroft Pressure Gauges | ASCO Valves | Katronic Flow Meters | KROHNE Flow Meters | VEGA Level Sensors | Rotronic Temperature & Humidity Sensors | SIKA Pressure Gauges

Technical Specifications

BrandSICK
CategoryGas Detection
Measured ValuesHCl, HF, NH3, H2O
Measurement PrinciplesDiode laser spectroscopy (TDLS)
Gas Flow Rate0 l/min … 20 l/min Depending on application
Measuring RangesHCl: 0 … 5 ppm / 0 … 3,000 ppm HF: 0 … 5 ppm / 0 … 2,000 ppm NH3: 0 … 25 ppm / 0 … 4,000 ppm
Dual Measuring RangesHCl / H2O: 0 … 5 ppm / 0 … 3,000 ppm (HCl) | 0 … 25 Vol.-% / 0 … 100 Vol.-% (H2O) NH3 / H2O: 0 … 25 ppm / 0 … 4,000 ppm (NH3) | 0 … 3 Vol.-% / 0 … 15 Vol.-% (H2O)
Response Time (t90)≤ 360 s Depending on sample gas flow
Accuracy≤ 2 % … ≤ 4 % Of measuring range full scale (depending on the application)
Sample Gas Temperature≤ +120 °C Temperature at analyser inlet
Process Pressure600 hPa … 1,200 hPa
Ambient Temperature0 °C … +50 °C
Storage Temperature–40 °C … +55 °C
Ambient Humidity≤ 85 % Non-condensing
Electrical SafetyCE
Enclosure RatingIP20
Analog Outputs3 outputs: 0/4 … 20 mA
Digital Outputs3 relay contacts: + 48 V AC/DC Pre-set for failure, maintenance and functional control
Digital Inputs2 potential-free contacts: / 24 V DC
Communication InterfaceSerial, PROFIBUS DP Remark: Only for HF Type of fieldbus integration: RS-232 Function: Proprietary service interface
Number1
OperationMenu-driven operation via LC-display and membrane keyboard
Dimensions (W x H x D)483 mm x 176.5 mm x 420 mm
Weight12.5 kg
Power SupplyVoltage: Standard version: 115 V / 230 V Frequency: 50 … 60 Hz Power consumption: Analyser: ≤ 500 W
Sample Gas ConnectionsSwagelok 6 mm: for 6/4 mm piping
Auxiliary Gas ConnectionsN2 flushing for oxygen measurement: Swagelok 6 mm: for 6/4 mm piping
Test FunctionsAutomatic check cycle for zero and span point